according to EN ISO 17665-1 standard and respect the
following conditions:
- Class B steam sterilisation in autoclaves conforming to
EN 13060 standard. Sterilisation time at least 3.5 minutes
at 135 °C.
3) Remove the bag from the autoclave.
-Storage
- Keep the handpiece in the sterilisation bag until it is
needed.
- Once sterilised, store the product in a dry and dust-free
place, if possible with a low microbial charge.
CAUTION!
Respect the validity date for the sterilisation!
40
Eng
Disposal
Sterilise the product and its accessories before
disposing of them. Refer the disposal of the product
to authorised companies qualified to dispose of
special industrial waste.
Technical assistance – Ivory direct line
HINT!
To request technical assistance, compile the on-line form
available at: www.ivory-dent.com