DANGEr!
Only fit the bur in the handpiece when the handpiece is
stopped. Never attempt to grip a bur while it is rotating,
even when it is nearly stopped.
DANGEr!
Never press the locking button on the handpiece while
the instrument is in use or has not quite stopped. Doing
so can cause the bur to become loose of cause the
locking system to overheat and break.
DANGEr!
AVOID CONTACT between the patient's cheek and the
locking button during use. This could lead to the bur
being accidentally released, resulting in burns or lesions
to the patient.
WArNING!
Never operate the handpiece without a bur in the spindle.
48
Eng
This could cause the handpiece to overheat, causing
damage or burns. Never exceed the maximum speed
specified by the bur manufacturer. Keep bur shafts clean
at all times; dirty shafts can cause dirt to enter the spindle,
leading to the bur slipping during rotation, or even
preventing the bur from fitting in the spindle correctly.
Never attempt to fit an instrument to a rotating micromotor.
DANGEr!
Never apply excessive pressure to burs. They could break,
bend or become difficult to extract.