- Cleaning the spray nozzles
Use the cleaning wire provided to gently remove dirt and
deposits from the spray nozzles. Clean and disinfect the
cleaning wire.
CAUTION!
Do not force the brush into the conduit. Damage to the
conduit can change the direction of spray with respect to
the bur and reduce the effectiveness of cooling.
- Cleaning the optic fibre glass (if fitted)
Clean the glass of the optic fibre with a pad damped in
alcohol. Remove all residual impurities.
- Lubrication
CAUTION!
Lubrication cycles:
- After every 15 minutes of rotation.
- Before every sterilisation.
CAUTION!
Lubricate only with Ivory oil OL101S.