Disconnect the electricity before filling or emptying the basin.
- The basin is to be filled with lukewarm tap water, with the addition of a cleaning solution
to increase cavitation.
- The operator is responsible for ensuring that the choice of cleaning agents is adequate for
the items to be cleaned.
- Before cleaning valuable items, you are recommended to carry out a small sample test;
then you can proceed with the remaining material.
-
Always use baskets or beakers; otherwise suspend the items.
-
Never allow the items to rest on the bottom of the basin.
- The basin must be cleaned regularly and the cleaning solution changed often to increase
cleaning action efficiency.
-
Do not reduce ventilation.
-
Do not use the ultrasound basin in a dusty environment.
Dust accumulation can interfere with regular apparatus operation.
- Carry out degassing every time the liquids are replaced, run the apparatus with the solu-
tion on its own for approximately 10-15 minutes.
- Having completed the instrument cleaning cycle, rinse it with drinking water (deminer-
alised water would be better in order to avoid stain formation) and dry it thoroughly. In
the case of medical instruments, carry out the sterilisation cycle.
IMPORTANT: all instruments, after being cleaned with the ultrasound equipment,
WARNING
IMPORTANT: always wear personal protection gear.
12
PART 4
OPERATING INSTRUCTIONS
OPERATION TIPS
must be sterilized before use.
EUROSONIC 4D
®