EUROSONIC 4D
®
1. Fill the basin with lukewarm tap water (max 60°).
2. Add a cleaning solution to the water.
3. After all the objects to clean (with their baskets/beakers) have been immersed in the solution, the
level of the liquid must be between the two notches that indicate the MIN and MAX level, visible
inside the basin.
4. Switch the unit on by pressing the POWER button on the back. The machine is set to run for 10
minutes at a temperature of 60°C. This is the standard setting every time the machine is switched
on. To change it follow the instructions in paragraphs 5) and 6).
5. Select the cleaning time desired using the keys next to the display TIME min. Press
and
to decrease. Holding the button down increases selection speed. The time setting can be
read on the display.6
6. Select the cleaning temperature desired, using the keys next to the display TEMP °C. Press
increase and
to decrease. Holding the button down increases selection speed. The temperature
setting can be read on the display.
7. Press START / STOP ULTRASONIC
display blink to indicate that the device is running.
8. Press START / STOP HEATING
to indicate that the heating function is on. While it is on, a small light on the display next to the
temperature, blinks when the resistor is heating.
9. At the end of the time setting, the ultrasound function shuts down automatically
IMPORTANT: switch off the heating function when the solution is below the minimum
WARNING
level (MIN mark on the basin)
To empty the basin use the tap on the back of the device. Turn the know of the tap counterclockwise to
drain the water.
IMPORTANT: do not drain the basin by tilting the device.
WARNING
This could interfere with the future proper operation and safety of the device.
DEGAS THE LIQUIDS EVERY TIME THEY ARE REPLACED
WARNING: AT THE END OF THE WORKING DAY, TURN THE MAIN POWER SWITCH OFF.
GENERAL FUNCTION
to start the cleaning cycle. The lights on the
to heat the solution. The lights on the display blink
to increase
to
13