pressure shocks. Insufficient opening of the shut-off valve could reduce actual flow
delivered.
•
Connect the accessory to the pressure outlet.
•
After completion of the therapy
•
Turn off at the shut-off valve by turning the hand wheel in a clockwise direction to stop
position. Do not use excessive force.
•
Vent gas pressure from downstream equipment.
•
Disconnect the accessory from the pressure outlet.
6c.2 AFTER USE
•
Turn off at the shut-off valve by turning the hand wheel in a clockwise direction to stop
position. Do not use excessive force.
•
Ensure that the flow control knob is on ZERO position (if any).
•
Disconnect all downstream equipment from user outlets.
•
Refit pressure outlet and flow outlet protection cap. Before refitting the caps, ensure they
are clean.
ACCESSORY
7
BED HANGER
•
Bed hanger is used with a guard. It is designed for suspension of medical package – cylinder
and combination valve with the guard – onto the patient's bed. The maximum hanger load is
with the use of a 5-litre cylinder.
GCE FILLING ADAPTORS
GCE recommends using GCE filling adaptors; Non GCE filling adaptors can damage
the non-return valve in the filling port and impact the safety of the product.
CLEANING
8
Remove dirt with a soft cloth damped in oil free soap water & rinse with clean water. Used
cleaning agents shall comply with oxygen cleanliness standards.
Disinfection can be carried out with an alcohol-based solution (spray or wipes).
If other cleaning solutions are used, check that they are not abrasive and they are compatible
with the product materials (including labels) and gas (convenient cleaning solution - i.e.
Meliseptol).
Do not use cleaning solutions containing ammonia!
Do not expose to water or any other liquid.
Do not expose to high temperature (such as autoclave).
To apply the cleaning solution do not spray it as the spray may enter into the inner parts
of combination valve and cause contamination or damage.
Do not use pressure wash as it could damage or contaminate the combination valve.
If the inner parts of the combination valve have been contaminated do not continue to
use the combination valve under any circumstances. It must be withdrawn from service.