Commissioning; Presettings - ProMinent DULCOTEST ICT 2 Instrucciones De Servicio

Ocultar thumbs Ver también para DULCOTEST ICT 2:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Commissioning

type of sensor
setting ?
Access to all setting menus can be blocked with an access code !
meas. parameter
setting ?
38
8
Commissioning
8.1

Presettings

Check first that the sensor is correctly connected to the
®
DULCOMETER
D1C (do not open the sensor).
type of sensor :
ICT 2
sensor data
are changed!
Check that the right sensor type has been selected (see "type of
sensor setting?")
mr: meas. range
cc: cellconstant
α: temperature-
coefficient
Set the measurement range of the sensor (see "meas. parameter
setting?". Bypass the second parameter set (press Enter))
IMPORTANT
Check settings in all menus when changing the range allocation.
mr :
20 mS/cm
cc :
1.120/cm
α :
2.10% °C
15 mS/cm 23.8 °C
mr 2 : 200 μS/cm
cc 2 : 1.120 /cm
2.10% /°C
α:
1106 μS/cm 25.0 °C