B. The product is small in volume, light in weight and convenient for carrying.
C. Low power consumption
2.2 Major Applications and Scope of Application
The Pulse Oximeter can be used in measuring the pulse oxygen saturation and pulse rate through finger.
The product is suitable for being used in family, hospital, oxygen bar, community healthcare, physical
care in sports (It can be used before or after doing sports and it is not recommended to use the device
during the process of exercising) and more.
The problem of overrating would emerge when the patient is suffering from toxicosis caused by
carbon monoxide, the device is not recommended to be used under this circumstance.
2.3 Environment Requirements
Storage Environment
a. Temperature : -40°C ~+60°C
b. Relative humidity : ≤95%
c. Atmospheric pressure : 500hPa~1060hPa
Operating Environment
a. Temperature: 10°C ~40°C
b. Relative Humidity : ≤75%
c. Atmospheric pressure: 700hPa~1060hPa
3. Principle
The principle of the Oximeter is as follows: An experience formula of data process is established taking
use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive Hemoglobin
(Hb) and Oxyhemoglobin (HbO2) in glow & near-infrared zones. The operation principle of the device is:
Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse
Scanning & Recording Technology, so that two beams of different wavelengths of lights can be focused
onto a human nail tip through perspective clamp finger-type sensor. The measured signal can be
obtained by a photosensitive element, information acquired through which will be shown on screen
through treatment in electronic circuits and microprocessor.
4. Technical Specifications
4.1 Main Performance
A. Display of SpO2 value, PR value, bar graph and pulse waveform.
B. Low-battery indication: low-battery indication appears when the battery voltage is too low to work.
Automatic standby function.
C. Adjustable screen brightness, the display direction can be changed automatically.
D. PR sound indication; sound prompt for over-limit, finger-out and low battery.
E. Memory function
F. The data can be uploaded to the terminal equipment by wired mode.
G. Charging function
H. Time display
4.2 Main Parameters
A. Measurement of SpO2
Measuring range: 0%~100%
Accuracy: 70% ~ 100%, ±2%; 0% ~ 69%, unspecified
B. Measurement of pulse rate
Measuring range: 30bpm~250bpm
Accuracy: ±2 bpm or ±2% (select larger)
C. Resolution
SpO2 : 1%, Pulse rate: 1bpm
D. Measurement Performance in weak filling condition
Figure 1