Record Menu interface as shown in Figure 7. It indicates that the device is storing when the red dot "R●"
in measurement interface flickers.
Short press the button to select the option to be adjusted, then long press the button to change the
value.
"Mode": record mode selection, including: "Auto" and "Manual" mode. Under "Manual" mode, select to
turn on / off memory by "Record".
Auto record: start recording after stable data appear, pull out the finger to finish recording a group of
data (99 group of data at most), the total duration does not exceed 72 hours.
Manual record: store up to 24-hour data.
When the memory is full, it will display "Memory is full!", then it will enter the standby mode after
several seconds. When exiting the standby mode, it will display "Memory is full!" to give user a prompt,
press the button again to enter the measurement interface.
Note: under manual mode, when "Record" is "ON", the device will prompt to clear the data stored last
time.
It will display "Recording" when there is no operation under record state for 30s, then it will enter
energy saving mode after several seconds, long press the button to exit this mode; short press the
button, it will display "Recording".
Note: under data recording state, after the display screen turns off automatically, in order to save
power, pulse sound indication will turn off automatically.
"Seg": data segment.
After setting, short press the button to select "Exit", long press the button to exit record menu and
return to main menu.
"Delete All": delete all records (auto record mode is shown as Figure 9).
Note: please upload data in time after recording, otherwise the data may be covered when the storage
space is full.
Note: the historical data will be deleted once switching the mode. Under recording state, the record
mode can not be switched; under manual mode, the record mode can be switched only when turning
off recording firstly.
C. Time setting
a. Connect the master device to synchronize device time
After connecting real-time transmission (refer to relative chapter (6.1.4) for the connection method),
select "Options"-"Synchronize Device Time" on PC software interface to synchronize the device time.
b. Set device time manually
Under main menu, short press the button to select "Clock", long press the button to enter its sub-menu
as shown in Figure 10.
Short press the button to select the option to be adjusted, then long press the button to change the
value. The device adopts 24-hour clock.
"set time": set the time, "yes": to set the time, "no": not set the time
"set year": set the year
"set month": set the month
"set day": set the day
"set hour": set the hour
"set minute": set the minute
Adjustable range for year: 2015 ~ 2045, month: 1 ~ 12, day: 1 ~ 30 (when there are 31 days in a month,
it is 1 ~ 31), hour: 1 ~ 23, minute: 1 ~ 59.
After setting, short press the button to select "Exit", then long press the button to exit time setting
interface and return to main menu.
Fig. 10 Time setting interface
Fig. 11 System menu interface
D. System setting and other options introduction
Under main menu, short press the button to select "System", then long press the button to enter the
interface as shown in Figure 11.
Short press the button to select the option to be adjusted, then long press the button to change the