7. Maintain, Transportation and Storage

7.1 Cleaning and Disinfecting

When using alcohol wipes to disinfect the device, please air dry or clean it with clean soft cloth.
7.2 Maintain
A. Please clean and disinfect the device before using according to the User Manual (7.1).
B. Please recharge the battery when the screen shows
C. Recharge the battery soon after the over-discharge. The device should be recharged every six
months when it is no regular used. It can extend the battery life following this guidance.

7.3 Transportation and Storage

A. The packed device can be transported by ordinary conveyance or according to transport contract.
The device can not be transported mixed with toxic, harmful, corrosive material.
B. The best storage environment of the device is -40 ºC ~ 60 ºC ambient temperature and not higher
than 95% relative humidity.

8. Troubleshooting

Trouble
The SpO2 and Pulse Rate can not
be displayed normally
The SpO2 and Pulse Rate are not
displayed stably
The device can not be turned on
The display suddeny turns off
The battery can not be full
charged even after 10 hours
charging time.
9. Key of Symbols
Symbol
Description
Refer to instruction
manual/booklet
%SpO2
The pulse oxygen
saturation(%)
bpm
Pulse rate (bpm)
Full-voltage
Two grid of power
Low-voltage
Close the alarm sound
indication
Possible Reason
1. The finger is not properly
positioned
2. The patient's SpO2 is too low
to be detected
1. The finger is not placed inside
deep enough.
2. The finger is shaking or the
patient is moving
1. The batteries are drained or
almost drained
2. Malfunction of the device
1. The product will enter standby
mode without signal for 5
seconds
2. The battery is drained away or
almost drained
The battery is broken
Symbol
.
Solution
1. Place the finger properly and
try again.
2. Try again; Go to a hospital for
a diagnosis if you are sure the
device works properly.
1. Place the finger properly and
try again.
2. Calm the patient down
1. Please recharge the battery
2. Please contact the local
service center
1. Turn on the device and insert
finger
2.Please recharge the battery
Please contact the local service
center
Description
1. The finger clip falls off ( no
finger inserted)]
2. Probe error
3. Signal inadequacy indicator
IP22
International Protection
WEEE (2002/96/EC)
This item is compliant with
Medical Device Directive
93/42/EEC of June 14, 1993, a
directive of the European
Economic Community.
One grid of power
Manufacturer
Date of manufacture