Do not use cleaning solutions containing ammonia!
Do not immerse the product in any liquid.
Do not expose to high temperature (such as autoclave).
Clean immediately after use.
Clean in position as shown in this chapter. Do not turn upside down.
Do not use spray to apply the cleaner, as spray can penetrate the inside
of the product and cause contamination or damage.
Do not use pressure water as it may damage or contaminate the product.
Do not use the product under any circumstances if the internal parts of
the product have been contaminated. It must be decommissioned.

8.2. ASSOCIATED EQUIPMENT

The used associated equipment shall comply with EN ISO 10079-3 and EN
1789 as amended and with appropriate Instruction for use:
Microbiological filter: Use a filter with the required connection.
Suction hose: Use suction hoses with an inner diameter of 6mm.
Collection container: Use the collection container with the required con-
nection.
Safety container: Use safety containers with the required connection
Before connecting any associated equipment or medical device to the
MediEject II, always check that they are fully compatible with connection
features & performances of the product.
9. MAINTENANCE
9.1. LIFE TIME
Life time and waste management
Maximum life time of the product is 10 years.
At the end of the product's life time (10 years maximum), the product must
be withdrawn from service. The provider of the device shall prevent the re-
use of the product and handle the product in compliance with "Directive of
European Parliament and Council 2006/12/ES of 5th April 2006 on waste".
There is a ventilation opening for the vacuum indica-
tor located below the face plate at the bottom of the
vacuum indicator. See the mark „-„
Keep the ventilation area clean and dry.
9/176
EN