Labels; Replacing The Blast Shield - COOK Medical Rhapsody H-30 Manual Del Usuario

Ocultar thumbs Ver también para Rhapsody H-30:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Labels

It is up to the user to maintain the safety labels and ensure that they are in good condition.

Replacing the Blast Shield

The blast shield is a replaceable optical window that is located between the output focusing lens
and the proximal surface of the fiber delivery system. This window protects the output focusing
lens from damage from debris that could be ejected from a damaged fiber.
The blast shield is located on a bracket that is inserted from the top of the output focusing lens
mount, just behind the laser aperture/fiber bulkhead connector.
Required Equipment: #2 Phillips-head screwdriver
Estimated Replacement Time: 1 minute
Procedure
Unplug the device from the power outlet before beginning this process.
1.
Unscrew and remove the vented decorative enclosure
strip at the top of the rear panel and set aside. (See two
screw locations above.)
Cook Medical © 2019
Screws
Return to Table of Contents
Return to Table of Contents
2.
Completely loosen the locking screw on the bayonet
latch. The locking screw is a captive screw, so it will not
fall out when completely loosened.
Chapter 6
Maintenance
Bayonet Latch
Locking Screw
1-41

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido