System Specifications

General Specifications
Electrical Requirements
Dimensions
Weight
Temperature (operational)
Temperature (storage)
Relative Humidity
Cooling
Electrical Protection
Type of Applied Part
Atmospheric Pressure
Degree of Protection
Laser Specifications
Laser Type
Wavelength
Power to Tissue
Energy/Pulse
Divergence
Application Mode
Pulse Duration
Repetition Rate
Delivery Device
Output Deviation
Aiming Beam
Cook Medical © 2019
HL-30A: 115 V ~ ±10%/50-60 Hz/15 A (20 W output)
HL-30B: 115 V ~ ±10%/50-60 Hz/20 A (30 W output)
HL-30E: 230 V~ ±10%/50-60 Hz/10 A (30 W output)
20 in (W) x 23 in (D) x 51 in (H)
50 cm x 59 cm x 129 cm
225 lbs
102 kg
50
°
F - 77
°
F,
10
C - 25
C
°
°
41
°
F - 104
°
F,
5
C - 40
C
°
°
From 30% to 85% (noncondensing conditions only)
Self-contained water-to-air heat exchanger
Class I
BF (Note: Optical Fibers are Type BF Applied Parts)
800-1060 hPa
IPX0 (IEC529) Not protected from solids. Not protected from
liquids. IPX8 for foot pedal.
Pulsed CTH:YAG (Ho:YAG)
2100 nm
up to 30 W
0.5 - 3.5 J
0.44 rad ±5%
Pulsed
~150 - 800 µs (short and long pulse - width modes)
5 - 20 Hz
Wide range of flexible silica fibers
Accuracy: The device contains an internal measuring system
to control the actual emission of the laser energy. The device
does not require user calibration.
Deviation: The laser energy values can diverge from the
preset values by a maximum of 20%.
Diode laser, green, 532 nm (adjustable < 5 mW)
Return to Table of Contents
Chapter 1
Introduction
1-8