INSTRUCTIONS).
N.B. The headgear (l) must not be disinfected.
8 - PRECAUTIONS
Do not clean the mask with solutions containing alcohol, aromatic
compounds, humidifiers, antibacterial agents or aromatic oils. Such
solutions would damage the product and reduce its life.
Prolonged exposure to direct sunlight or heat may cause damage
to the mask.
If any visible damage is noted (cracks, tears, etc.), the damaged part
must be replaced.
Check carefully to make sure there are no soap deposits or foreign
bodies inside the mask.
Check regularly that the vents are not obstructed. Clean with a soft
brush if necessary.
9 - HOW TO REASSEMBLE THE MASK
Insert the anti-suffocation fitting (h) in the main body (b) of the mask,
pressing gently.
Remount the forehead rest (d) on the forehead support (c), keeping
the curved edge facing upwards.
Insert the swivel fitting (i) in the anti-suffocation fitting (h), pressing
gently.
Fit the grooved side of the cushion (a) into the frame of the main
body (b). Fit on the top first, then the sides of the mask, pressing
the cushion in until the whole edge is up against the six stops on the
main body (Fig. D).
Fit on the cap (g).
Pass the two lower straps of the headgear (l) through the slots in
the clips (m).
Attach the two upper straps of the headgear (l) to the main body (b).
Clip the two clips (m) back onto the main body (b).
Turn the adjusting knob (e) to the reference position you noted
previously.
ENGLISH
10 - PRESSURE/OXYGEN PORT
If additional oxygen or a pressure reading is required, use the port
in the main body (b).
Open the cap (g) by lifting upwards, without removing it.
Push the pressure/oxygen tube firmly into the port.
11 - TROUBLE-SHOOTING CHART
PROBLEM
• No air or
insufficient air
comes out of
the mask
• The mask is
noisy
• The mask does
not fit the face
snugly
• Exhaled air does
not flow out
and difficulty
inhaling

12 - DISPOSAL INSTRUCTIONS

The components of the mask are made of the following materials
and must be disposed of in accordance with the applicable laws.
13
POSSIBLE CAUSES SOLUTIONS
• Flow generator off
• Switch on the flow
generator
• Flow generator
• Regulate the flow
regulated wrongly
generator
• Patient circuit tube
• Connect the tube to
not inserted
the flow generator
and mask
• Cushion mounted
• Mount the cushion
incorrectly and air
correctly
leak
• Anti-suffocation
• Clean or replace
fitting dirty or
the anti-suffocation
damaged
fitting
• Mask applied
• Reposition the
incorrectly
headgear
• Vent ports
• Remove the
obstructed
obstruction