Lektrafuse HF Generator GN200
7.4

Failure messages

A failure is indicated as follows:
Clear sound sequence (as necessary, may interrupt the sound of the HF
discharge)
The red error light 6 on the front side of the HF device flashes
Failure report in the display 5 on the front side of the HF device
Failure code
Cause
F001
Software error on the HF device
F002
Dosage error of HF device
F003
Hardware error
F004
Internal communication error of the HF device
F005
Overheating of HF device
F006
Error on the RF on/off button 2 (activated when switching on) Switch off HF device, wait 5 s, release RF on/off button 2,
F007
Error with HF energy output
F008
Fault: HF outside the calibration
F009
Fault: Internal power supply
7.5

Problem solving

Finding
HF device does not switch on
The device is switched on, but no displays are illumi-
nated and the self-test is not being carried out
Self-test fails
HF device is active ("power ON signal lamp 9" light
up), and accessories are connected, but there is not HF
discharge
16
®
Remedy
Ensure that the power cord has been correctly plugged into the power supply 13 of the HF
device.
Ensure that the power cord is connected to a functioning power source.
Ensure that the power OFF/ONN switch 12 of the HF device is in the power ON position.
Check the fuse in the fuse box and replace if necessary, see Fuse replacement.
If the error continues, replace power cord.
Switch off HF device, wait 5 s, and switch HF device on again.
If the error continues, contact your national B. Braun/Aesculap agency, see Technical Ser-
vice.
Switch off HF device, wait 5 s, and switch HF device on again.
If the error continues, contact your national B. Braun/Aesculap agency, see Technical Ser-
vice.
Verify that the required foot switch (if used ) is properly connected and plugged in.
Ensure that the Caiman Seal & Cut Instrument is correctly connected and plugged in (green
ring 1 is lit up).
In order to start the HF discharge, activate the "RF on/off" button 2.
If the error continues, replace Lektrafuse instrument.
Note
If the failure occurs during HF discharge, the vessel sealing may be incom-
plete or insufficient.
In order to ensure that sufficient hemostasis has been achieved, follow the
procedure for regrasp errors, see Regrasp error.
Note
If the failure continues after turning the device off and on again twice, con-
tact your national B. Braun/Aesculap agency, see Technical Service.
Remedy
Switch off HF device, wait 5 s, switch HF device on
Switch off HF device, wait 5 s, switch HF device on
Switch off HF device, wait 5 s, switch HF device on
Switch off HF device, wait 5 s, switch HF device on
Switch off HF device and leave to cool
Check the air stream around the generator
switch on HF device
Switch off HF device, wait 5 s, switch HF device on
Switch off HF device, wait 5 s, switch HF device on
Switch off HF device, wait 5 s, switch HF device on