Acoustic Warning Signal; Monitoring Functions - B.Braun Aesculap GN200 Instrucciones De Manejo

Ocultar thumbs Ver también para Aesculap GN200:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 58
Aesculap
Lektrafuse HF Generator GN200
2.5

Acoustic warning signal

Status
System failure
REGRASP error
System error
System warning
Start HF discharge
HF discharge active
HF discharge ended
Power up self-test
2.6

Monitoring functions

Self-test
As soon as it is switched on, the unit performs a self-test of the control
elements, the acoustic warning signal, the microprocessor and the hard-
ware function. During this phase, the message "Performing Self Test"
appears in the display 5.
Fig. 1
Continuous test cycle during operation
During operation, safety-relevant functions and signals are monitored
through a continuous test cycle. As soon as a critical error is detected, the
HF device terminates the HF activation. In display 5, the error number of
the corresponding error is displayed and an acoustic warning signal is
emitted, see Troubleshooting list.
6
®
Acoustic warning signal
3 repetitions of: 1 long beep 3 KHz, 1 beep 2 KHz Following each system failure, if identified (F001,...)
3 repetitions of: 1 beep 2 KHz,
1 beep 3 KHz
3 beeps, 3 KHz
3 beeps, 3 KHz
2 beeps, 480 Hz
Repeating beep 480 Hz
3 beeps, 530 Hz
1 beep 2 KHz
Notes
Following each regrasp error, if identified
Following each failure report, if identified (E001,...)
Following each warning, if identified (W001,...)
When starting the HF discharge
Continually during HF discharge
When a sealing process has been successfully completed (no
failure reports or errors)
During the self-test, in order to check that the signal tone is
functioning

Publicidad

Tabla de contenido
loading

Tabla de contenido