7 Error limits

Assembly
a.
6.2.4 Air vent cap for micro filter with Luer-cone
For sterile media we recommend the air vent cap with Luer-cone to attach a micro filter. This provides
increased protection against contamination by displacement air (Accessories/spare parts, p. 66).
Assembly
a.
c.
7 Error limits
20 °C
!
Ex
Type Digital • Easy Calibration is
manufactured under U.S. Patent
5,957,330.
Error limits
Nominal volume ml
    1
    2
    5
  10
  25
  50
100
52
Operating manual
a.
Place the PTFE sealing ring on the bottle thread or the
screwed-on adapter and screw the instrument onto the
bottle.
a.
Unscrew the air vent cap (see 'Assembly Drying tube', Fig.
a.).
b.
Screw in the air vent cap with a Luer cone.
c.
Place the PTFE sealing ring on the bottle thread or the
screwed-on adapter and screw the instrument onto the
bottle.
d.
Insert a commercially available sterile filter into the Luer
cone.
Error limits related to the nominal capacity (= maximum vol-
ume) indicated on the instrument, obtained when instru-
ment and distilled water are equilibrated at ambient temper-
ature (20 °C/68 °F). Testing takes place according DIN EN ISO
8655-6 with a completely filled instrument and with uniform
and smooth dispensing.
μl
A*
± %
0.6
    6
0.5
  10
0.5
  25
0.5
  50
0.5
125
0.5
250
0.5
500
μl
CV*
%
0.2
    2
0.1
    2
0.1
    5
0.1
  10
0.1
  25
0.1
  50
0.1
100
9974 90