6 Operation
f.

6.2 Accessories

6.2.1 Flexible discharge tube with recirculation valve
For serial dispensing the flexible discharge tube can be used for the bottle-top dispenser Dispensette
S and Dispensette
The specified accuracy and coefficient of variation of the instrument are only obtained for volumes >
2 ml and by gently approaching the upper and lower stops. The coil of the tubing can be stretched to a
maximum length of 800 mm. Before use, the tube must lie in regular loops and must not be twisted.
The operating exclusions for the respective instrument in use apply.
Assembly
Use only undamaged tubing
There should be no visible damage to the discharge tube (e.g. kinks, etc.). Each time you
are going to use the tubing, examine it carefully.
> To dispense aggressive liquids, you should take safety measures in addition to the
normal precautions. We recommend use of a protective shield.
> The bottle must be supported using a bottle stand.
> To help avoid reagent splashing from the tube, always grip the tube firmly by the
handle and replace into the holder after use.
> For cleaning rinse the tube carefully.
> Do not dismantle!
1
2
50
Operating manual
®
S Organic (Accessories/spare parts, p. 66).
After using the piston, always press it down to the lower
stop (parking position). If the piston is not pressed down to
the lower stop, unintentional media leaks may occur.
The filled status of the instrument must be specially
marked during cleaning.
WARNING
Prerequisite:
>
If the instrument was in use, it must be cleaned before in-
stalling the flexible discharge tube (Cleaning, Clean-
ing, p. 56).
a.
Instruments with a recirculating valve should be set to
'Recirculate', and the valve lever pulled upwards to re-
move it.
b.
Slide the discharge tube housing all the way up, then pull
it forward with gentle up and down motions (Fig. 1).
c.
Push the flexible discharge tube holder from the bottom
of the valve block and tighten it (Fig. 2). In doing so, the
instrument must not be mounted to a bottle. Install the
receiver tube.
d.
Press the plug of the recirculation valve downwards.
e.
Slide the flexible discharge tube housing into the valve
block up to the stop (Fig. 3).
NOTICE
NOTICE
®
9974 90