Chapter 1, General Description - MSA Altair QuickCheck Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Chapter 1,
General Description
The Altair QuickCheck Station is designed to
test the Altair and Altair Pro Single Gas
Detectors, verifying:
instrument response to gas (bump check)
alarm functionality.
The Test Station:
uses a microphone
to detect the audible alarm
electronically checks the instrument's
visual LED alarm and
standard vibrating alarm
tests the instrument's response to a
known concentration of gas
is compatible with:
all three Altair Detector versions:
CO, H 2 S and O 2
Altair Pro Detector versions:
O 2 , CO, CO Steel, CO Fire,
H 2 S, HCN, Cl 2 ,
ClO 2 , SO 2 , NO 2 and NH 3
has three LED indicators:
Alarm Test LED
Gas Test LED
Pass/Fail and Confidence Pulse LED
(indicates unit is receiving power).
NOTE: The Altair and Altair Pro instruments
record the date, time, and result of a bump
test internally in their event recorder:
25 events are stored in the
Altair Detector
50 events are stored in the
Altair Pro Detector.
1
4
5 6
7
ITEM
NUMBER DESCRIPTION
1
REGULATOR (sold separately - manual version)
2
GAS CYLINDER (sold separately)
3
ALTAIR INSTRUMENT (sold separately)
4
TEST STATION HOUSING
5
ALARM TEST LED
6
GAS TEST LED
7
PASS/FAIL LED
Figure 1-1. Altair QuickCheck Station
Components
2
3
1-1

Publicidad

Tabla de contenido
loading

Tabla de contenido