Test Station Set-up

1. Place the unit:
on a level surface
in an atmosphere with an ambient noise
level of 100 dB or less.
2. Test station power requirements are:
115-230 VAC 50-60 Hz input
(MSA power supply)
6 V output to the test station.
3. Connect power cord into the receptor
at the rear of the unit.
4. To connect tubing, attach:
one end of tubing
to the regulator barb fitting
other end of tubing
to the test station barb fitting.
5. Connect regulator to the necessary
gas cylinder.
See TABLE 1-2 for necessary gas
cylinder concentrations.
Only the specified gas concentrations
may be used with the Altair QuickCheck
Station.
Table 1-2.
Necessary Gas Cylinder Concentrations
GAS
ALTAIR /
CONCENTRATION
ALTAIR PRO
USED FOR
SENSORS
QUICKCHECK
H 2 S
40 ppm
40 ppm
CO
60 ppm
O 2
5%
NO 2
10 ppm
10 ppm
SO 2
10 ppm
10 ppm
NH 3
25 ppm
25 ppm
CL 2
10 ppm
10 ppm
CLO 2
2 ppm Cl 2
2 ppm Cl 2
HCN
10 ppm
10 ppm
• All Cylinders are: RP - 58 LTRS, 500 PSI
EC - 34 LTRS, 500 PSI
NOTE: Required gas concentrations
are also listed on the bottom of the test station.
CYLINDER
PART
TYPE
NO.
Econo-Cal
711062
RP
467897
100 LTRS Steel RP
710882
100 LTRS Steel RP
493580
Econo-Cal
711068
RP
808977
Econo-Cal
711070
RP
808978
Econo-Cal
711078
RP
814866
Econo-Cal
711066
RP
806740
Econo-Cal
711082
RP
10028080
Econo-Cal
711072
RP
809351
1-3