Chapter 4,
Specifications
Table 4-1. Specifications
FOR USE WITH
TESTS PERFORMED
VISUAL INDICATORS
VERSIONS
AVAILABLE
TESTING TIME
DIMENSIONS
WEIGHT
POWER SUPPLY
REQUIREMENTS
OPERATING
TEMPERATURE
ALTAIR
SINGLE-GAS DETECTOR
ALTAIR PRO
SINGLE-GAS DETECTOR
LED INDICATORS
FLASHING AMBER
SOLID RED
SOLID GREEN
CHECKMARK
MANUAL
AUTOMATIC
AVERAGE
All versions
O 2 , H 2 S, CO, CO Fire, CO Steel, SO 2 , NO 2 , CL 2 , CLO 2 , NH 3 , and HCN
Visual alarm activation
Vibrating alarm activation
Audible alarm activation and verification
Sensor response to gas
General pass fail indication
FOR ALARM TEST, GAS TEST, AND GENERAL PASS/FAIL
Test in in progress
Failed Test
Passed Test
Displays on ALTAIR instrument for 24-hours following a passed test
Features a .25 lpm manual regulator that user turns ON during testing
Features a .25 lpm automatic regulator that pulls gas as it is needed
Less than 30 seconds
2" h x 4-¾ " w x 6-½ " l (5.1 cm x 12.1 cm x 16.5 cm)
8.6 oz. (245 g)
115/230 VAC, 50/60 Hz.
32º F to +104ºF (0ºC to +40ºC)
4-1