6.2.2.4.1 REPLACE LAMP NOW
The Replace Lamp Soon message is an information warning and does not require any
action from the user. This message displays whenever the currently installed lamp has
reached 950 hours of usage. This message begins to flash whenever the lamp reaches
975 hours of usage. This warning is shown because the lamp will no longer turn on when
the lamps usage reaches 1000 hours.
6.2.2.4.2 CHANGE FILTER AND RESET
The Change Filter and Reset message is displayed flashing whenever the unit runtime
hours an increment of 250 hours (250, 500, 750, 1000, 1250, etc.).Every 250 hours the
unit filter is supposed to be replaced in order to keep the unit, and the lamp, cool enough
to avoid damage. After replacing the filter, the user must go into the Filter Reset menu and
press the Reset Filter button in order to clear this warning.
6.2.2.4.3
This screen will advise if the air filter is due for replacement. Air filter replacement every
250 hours of unit operation isadvised. When air filter replacement is due there will be a
flashing signal to the left of
LIT-218 Sunoptic Surgical
Rev. C
Replace Lamp Soon
Lamp Hours :
_
100
MENU
Lamp Hours :
_
100
MENU

RESET FILTER

Lamp Hours :
MENU
RESET
FILTER.
®
Page 16 of 170
950
+
STANDBY
Intensity Adjust
Change Filter and Reset
950
+
STANDBY
Intensity Adjust
250
RESET FILTER
Reset Filter
(English)