6.2.2.4.1
This screen will advise if the air filter is due for replacement. Air filter
replacement every 250 hours of unit operation isadvised. When air filter
replacement is due there will be a flashing signal to the left of
After installing a new filter, touch
the next 250 hours of unit operation.
6.2.2.4.2
The error log functiondisplays error codes useful for repair service. The minus
plus
controls will cycle through the error codes logged.