Supplement to the instructions for use
Keyswitch
The Movita lift and Forta lift ceiling supply units are
set up for operation with the following Dräger
devices if they are equipped with the keyswitch
shown below:
– Perseus anesthesia workstation
– Other Dräger devices, e.g., Primus anesthesia
workstation or Mova Cart equipment cart
On the media column there is a keyswitch which
can be set to the following positions:
Perseus
Other
Dräger
devices
C
A Operation of Perseus
Make sure that the keyswitch on the ceiling
supply unit is in position (A). Contact service
personnel if necessary.
B Docking and undocking of Perseus
This switch position may only be used by
service personnel when docking or undocking
Perseus. Operation in this switch position is
neither possible nor permissible, as safety
functions of the ceiling supply unit are switched
off in this position.
C Docking, undocking, and operation of other
Dräger devices, e.g., Primus or Mova Cart
6
Perseus
A
B
Docking, undocking, and operation of Perseus
are neither possible nor permissible in this
switch position as the Perseus safety sensor is
not operational in this position.
Docking Perseus to the ceiling supply unit
Only service personnel may dock Perseus to the
ceiling supply unit or undock it. When docking or
undocking, follow the instructions for use of the
ceiling supply unit, the instructions for use of
Perseus, and the associated supplements.
Connect the cable of the remote control to the
socket on the ceiling supply unit to enable the
functions of the remote control and the safety
sensor.
WARNING
Risk of personal injury
If the ceiling supply unit has not been set up
for operation with Perseus or the keyswitch is
in an incorrect setting position, the safety
sensor will be inoperative.
Only operate Perseus with ceiling supply
units that have been set up for Perseus and
have the keyswitch in the correct setting
position. If necessary, contact service
personnel for clarification and experts if the
keyswitch is missing.