8.3. Beginning therapy

• Connect the cable to the power supply, ensuring in advance that the voltage
matches the details given on the specification plate.
• The light unit can be switched on or off with the START/STOP switch. Even if the
fluorescent tubes are switched off at the switch, the BiliBed remains operational,
i.e. connected to the power supply. The equipment is only completely switched off
when the plug is disconnected from the power supply.
Note
The START/STOP switch can be omitted entirely. The light unit can be started and
stopped by plugging in or unplugging the power cord to the unit.
Caution
The therapy blanket must be positioned correctly to ensure successful treatment. The
baby must lie above the therapy lamp. The position of its head and feet must be
according to the drawing on the light unit.
Caution
Supervise the infant during therapy. Regular checks are necessary:
• Check the bilirubin levels. Photoisomeres of bilirubin may cause toxic effects.
• Phototherapy influences the baby's temperature. A regular check is necessary.
• Phototherapy influences the infant's water balance. In order to prevent dehydration,
it may be necessary to feed the baby more often.
Attention
Auxiliary devices which are used in the same area as the BiliBed must comply with the
general requirements for safety according to IEC 60601-1.
EN
11