Using the Configuration Page
Certain set-up and configuration options are available on the password-protected
3M™ Attest™ Auto-reader 390 Web App Configuration page. Ideally, the 3M™ Attest™
Auto-reader 390 should be empty (i.e., have no incubating BIs) when it is reconfigured.
Navigate to the Configuration page by entering either:
1) http://attestxxxxxx/protect/config.htm (where xxxxxx is the 6-digit unit serial number), or
2) http://xxx.xxx.xxx.xxx/protect/config.htm (where xxx.xxx.xxx.xxx is the IP address)
Click on the desired radio buttons to change settings.
Auto-reader Type
Select the configuration desired for wells 1-5/6-10 by clicking on the appropriate radio button.
Language
Selected setting indicates language in which text on 3M™ Attest™ Auto-reader 390 Web App pages will appear on the computer screen.
Selecting Enable
Pushbutton
Configuration
Selecting Disable
Selecting Enable
Buzzer
Configuration
Selecting Disable
Attest Auto-reader
Naming the 3M™ Attest™ Auto-reader 390 may be useful in facilities with multiple units. The assigned nickname will appear on the 3M™ Attest™
390 Nickname
Auto-reader 390 Web App Status page.
Default: Setting is checked indicating the Web App will use the Dynamic IP address automatically assigned when the 3M™ Attest™ Auto-reader 390 is
Enable DHCP
connected to the network. Remove the check if you wish to use a Static IP address assigned by your facility IT.
IP Address
Enter Static IP address assigned by IT.
Gateway
Enter Gateway.
Subnet Mask
Enter Subnet Mask.
Primary DNS
Enter Primary DNS.
Secondary DNS
Enter Secondary DNS.
MAC Address
XX:XX:XX:XX:XX:XX (Unique MAC address assigned to device. This value cannot be changed.)
Use Computer
Check this box if you wish to set the clock on-board the 3M™ Attest™ Auto-reader 390 using the computer's clock.
Time
Attest Auto-reader
Alternatively, leave the Use Computer Time box unchecked and complete the date and time boxes to set the 3M™ Attest™ Auto-reader 390 clock.
390 Time
After changing desired settings, click on Save Config button to save your changes. A new screen confirming the settings were successfully changed appears.
allows the user to configure the 3M™ Attest™ Auto-reader 390 from the buttons on the unit itself.
prevents the user from being able to configure the 3M™ Attest™ Auto-reader 390 from the buttons on the unit itself.
enables the audible alarm. The speaker icon will appear on the 3M™ Attest™ Auto-reader 390 LCD panel.
disables the audible alarm. The speaker icon will disappear from the 3M™ Attest™ Auto-reader 390 LCD panel.
in the web browser address bar and pressing Enter. A security screen will appear. Fill in
the fields as follows:
User name: manager
Password: Attest
and press OK.