2 Description

This generator will produce a constant stream of ultra high purity hydrogen at a predetermined flow rate and pressure when connected to a
suitable power supply and fed with a suitable quality of deionised water. It is suitable for use in laboratories and light industrial environments and
is non-hazardous for transportation purposes.
In order to guarantee the optimum efficiency of the PEM cell, this generator must be installed and running within three months of dispatch
from Parker Hannifin. Failure to do this may invalidate the warranty.
If the generator has been stored for a period of 90 days or more, it will automatically perform a 240 minute (4 hours) initialisation sequence on
power up. This sequence, which cannot be aborted, is necessary to guarantee the correct hydration of the cell and operation of the micro-
dryer.

2.1 Technical Specification

This specification is valid when the equipment is located, installed, operated, and maintained as specified within this user guide.
Water
Water quality
1
Consumption (Approximate)
2
Supply pressure (max)
2
Supply flow rate (max)
2
Supply temperature (max)
Ultra High Purity Hydrogen (H
Outlet flow rate
Outlet pressure
3
Purity
Purge outlet
Purge flow rate (@ 7.7bar g)
Mechanical Connections
Hydrogen outlet
Water drain
Automatic water fill inlet
(factory or field fit option)
Overflow drain
Purge outlet
Spillage drain
Electrical Data
Connection type
Supply voltage range
Power consumption
4
Fuse
1.
Based on full flow with 24 hour 7 day operation at 22
2.
Applies to generators with auto water fill only.
3.
The balance is O
and moisture.
2
4.
Anti Surge (T), 250V, 5 x 20mm HBC, Breaking Capacity 1500A @ 250V, IEC 60127, UL R/C Fuse
Units
20H-MD
L / week
1.69
bar g
(psi g)
L/min
o
C
o
(
F)
)
2
ml / min
160
bar g
(psi g)
%
ml/min
Vac
W
170
A
o
o
C (72
F) ambient temperature.
40H-MD
Deionised, ASTM II, >1 M , <1 S, filtered to <100 m
2.41
0.1
1.45
20
(68)
250
0.69 - 6.89 ± 0.05
(10 - 100 ± 0.73)
> 99.9999%
48.125 - 57.75
1»8Ý Compression fitting
Quick release push in fitting
Blank
Quick release push in fitting
1»4Ý Compression fitting
(DO NOT CONNECT – Service use only)
1/2Ý Barbed push on fitting
IEC320
100 - 230v 50»60 Hz
230
5
60H-MD
4.82
1
500
96.25 - 115.5
Quick release push in
320
110H-MD
9.64
1100
fitting
640
5 (x2)
3

Publicidad