11. Start the sterilization cycle up pressing the
The autoclave is starting the preheating phase (on the display you can read HEATING).
When the machine is reaching the necessary pressure for the chosen cycle, the sterilization
phase is starting (on the display you can read "STERILIZ").
On the display is shown the remaining sterilization time.
When the sterilization cycle is completed, the machine is discharging the pressure and the
drying phase is starting (on the display you can read "DRYING").
On the display is shown the remaining drying time.
When the sterilization cycle is completed on the display you can read "END".
IMPORTANT: BEFORE OPENING THE DOOR BE SURE THAT THE CHAMBER IS NOT
UNDER PRESSURE. DO NOT OPEN THE DOOR IF THE PRESSURE VALUE SHOWN
ON THE DISPLAY IS HIGHER THAN 0,03 BAR
12. Open the door
1A
REMARK 1: the closing bolt is complete with a notch to not allow the door opening if the
chamber is under pressure. To make the opening easier push the door toward the chamber;
REMARK 2: if you do not open the sterilization chamber immediately after the end of the
sterilization cycle the steam is condensing on the door, so drops will drop while opening the
door.
IMPORTANT: TO AVOID BURNINGS KEEP AT THE RIGHT DISTANCE WHILE
OPENING THE DOOR AND KEEP SAFE FROM STEAM LOSSES
13. Keep the instrument from the sterilization chamber using protection gloves the special gripper.
after that you have checked that the chamber is not under pressure.
button.
START/STOP