11. MAINTENANCE
CAUTION: DISCONNECT ALWAYS THE MAIN SUPPLY BEFORE EACH TECHNICAL
INTERVENTION!
DAILY ROUTINE MAINTENANCE
11.1.
•
Clean the seal on the door
the gasket using a wet cloth or the soft part of the supplied sponge.
This cleaning operation must be carried out to remove possible dirt which can cause a
pressure loss in the sterilisation chamber and a possible seal cut.
•
Check the water level in the tank (see paragraph 8).
WEEKLY ROUTINE MAINTENANCE
11.2.
•
To remove the deposits on the chamber floor, clean with the abrasive part of the supplied
sponge after removing the tray-holder.
To moisten the sponge, only use distilled or demineralized water.
•
Empty the water tank (see paragraph.9).
QUARTERLY MAINTENANCE
11.3.
•
Lubricate the hinges and the closing pin with silicon oil.
•
Adjusting the door
If the handle
Photo F to increase the pressure of the gasket on the chamber.
and on the outer edge of the sterilisation chamber sealing
1A
:
1A
is too frail while closing and blocking the door, turn it counterclockwise
2A