7. Service
7.1 Spare parts, accessories and consumables
Description
USB-Adapter Mini-USB to USB A
NANOCOLOR
®
USB T-Set
Adaptor cable for use in car
7.2 Error messages
The device shows error messages in case of problems. The reason for those messages
can be due to improper handling or faults in the device. In case of recurrent errors or
limited functionality of the instrument, please contact MACHEREY-NAGEL.
Error message
"Digestion interrupted"
7.3 Disposal
In compliance with the local and national legal regulations (EU Directive 2012/19/
EU), MACHEREY-NAGEL disposes old instruments.
Note: Disposal using public waste disposal facilities is not permitted. In case of disposal,
please contact your MACHEREY-NAGEL representative.
7.4 Cleaning the device
The instruments housing can be cleaned with a wet cloth. Splashes and spills need to be
immediately removed from the instrument. Generally, the heating block should be kept
clean. When contaminations inside the instruments occur which are caused by leakage
of a tube, please contact MACHEREY-NAGEL or your local distributor.
Cause
An external USB device has
to be connected to run the
chosen option; the connected
USB device was not reco-
gnized.
A connection to a
NANOCOLOR
USB T-Set
®
is required for this option; the
connected NANOCOLOR
USB T-Set was not reco-
gnized.
The temperature during the
heating process deviates by
> 4 °C to the set temperature.
The power connection was
interrupted during the heating
process.
NANOCOLOR
VARIO Mini V 1.0 / 01.16
®
919937
919921
919938
Solution
Please connect a USB
device or check the correct
connection of the USB
device.
Please connect a
NANOCOLOR
USB T-Set or check the
®
correct connection of the
NANOCOLOR
Do not insert cold tubes
during the heating process.
Check the power supply.
REF
®
®
USB T-Set.
31