REVISION V1.6
PRINTED 03/2017
7.3.2 Manual cleaning (see 7.1.2)
7.3.3 Rinsing off the cleaning solution (see 7.1.3)
7.3.4 Preparing the cleaning and disinfecting
device for endoscopes:
Place the cleaned endoscope in the basket of the
cleaning and disinfecting device for endoscopes in
accordance with the manufacturer's instructions and
connect the endoscope to the relevant system if
required. Place the accessories (e.g. valves, cleaning
brushes) in the accessories compartment. Then push
the basket into the cleaning and disinfecting device
for endoscopes, close the door, select the program
and start it.
7.3.5 Removing the endoscope from the cleaning
and disinfecting device for endoscopes:
Remove the endoscope with disinfected hands or
sterile disposable gloves and carry out a functional
test (see 5.). The endoscope can then be used for
patient examinations once more.
For storage or preservation, the endoscope must be
stored completely dry, protected from dust, and
preferably hanging in a special endoscope cabinet.
Store the vent cap dust-free and dry and do not insert
it into the endoscope during storage.
8. Hydrogen peroxide sterilisation
(STERRAD® method)
The effectiveness of the sterilisation of the flexible
nasopharyngoscope was verified with STERRAD
100S (endoscope double-packed in ASP Tyvek®
bags). The material compatibility is provided using
STERRAD 100 S and STERRAD 100 NX DUO
Cycle. Observe specifications of the manufacturer
(ASP) regarding the corresponding method.
9. ETO gas sterilisation
ETO
gas
sterilisation
FENTEXmedical nasopharyngoscope in respect of
material compatibility. The process parameters must
be performed in accordance with the information of
the ETO gas sterilisation manufacturer. Desorption
times prescribed according to the gas concentration
must be observed.
10. Electrical safety
The degree of electrical insulation is fixed by the
manufacturer
of
the
FENTEXmedical endoscope and its accessories.
Many
years
of
experience
instruments exclusively manufactured and repaired
by FENTEXmedical GmbH do not give rise to any
safety risks.
11. Service
700700FX / 700710FX
is
suitable
for
equipment
used
with
have
shown
CO#04-0127
FENTEX
mechanical parts to any outside company, therefore
only Fentex Germany, or its subsidiaries are able to
repair Fentex medical endoscopes with original parts
and
to
specification and safety of the product.
Additionally, the FENTEX warranty for the medical
endoscope is void if a repair is done by an
unauthorised repair shop. Then Fentex is no longer
responsible for the original specification and for the
safety standards of the product.
The vent cap must be connected during transport in
the crate, as otherwise the endoscope may be
damaged by pressure changes.
the
a
that
12 / 24
does
not
supply
their
guarantee
the
original
info@fentexmedical.com
optical
and
technical