Integra MAYFIELD XR2 A2079 Manual De Instrucciones página 14

Sistema radiotransparente
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 82
Optional Automatic Wash Cycle:
Any deviation from this guideline could result in damage to equipment as well as
improper cleaning results. Parameters given below must be followed exactly in order
to assure proper cleaning and avoid damage to the equipment.
Disassemble the device as indicated in section Cleaning of the MAYFIELD Infinity XR2 Base
Unit.
Rinse components under warm tap water prior to placing in washer.
Arrange device in a way to avoid contact between components.
Select instrument cycle and ensure proper programming.
Table 1 Instrument Cycle
Phase
Pre-Wash 1
Enzyme Wash
Wash 1
Rinse 1
Thermal Rinse**
* Hot Tap water not to exceed 82.2C
** Optional phase for disinfection of components
Remove from washer and dry completely if needed.
Inspect all components to ensure there is no visible organic debris or residue from cleaning
agent. Repeat process if any visible debris or residue is detected.
CAUTION!
Time in minutes
04:00
04:00
10:00
00:30
02:00
Table 1 Instrument Cycle
Page 12
Water
Detergent type &
Temperature
Concentration
Cold tap H
O
2
Endozime® AW
Hot tap H
O*
2
A.P.A., 1:128 ratio
Renu-Klenz™,
60° C
Hot tap H
O*
2
82.2° C
N/A
Triple plus w/
1:256 ratio
N/A
N/A

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Mayfield xr2 a2079e

Tabla de contenido