6.3 Operating mode and range
The autoclave allows the configurations of the sterilisation cycle in
accordance with the features to be sterilised:
Program type
Solids atmospheric purge:
Solids two fraction purge:
Solids three fraction purge:
Liquids atmospheric purge:
Sterilisation time:
Drying time:
6.4 Pre-established cycles
The Autester ST DRY PV II have 10 preset programs which can be modified
depending on the users requirements.
Program
0
Not wrapped instruments
1
Wrapped instruments
2
Prions
3
Not wrapped delicate
4
Wrapped delicate
5
Instruments
6
Disinfection
7
Liquids
8
Solids
9
Solids
* Purge types:
1. Atmospheric purge
2. Two phase vacuum purge
3. Three phase vacuum purge
6.5 Test programs
The Autester ST DRY PV II have two test programs:
Bowie Dick test
Vacuum test
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta @jpselecta.es - website: http://www.jpselecta.es
With external
internal sensor
105-134ºC
115-134ºC
115-134ºC
105-134ºC
from 3 to 59 minutes
from 0 to 99 minutes
Sterilisation Drying
Temperature
Time
ºC
(minutes)
134
4
134
12
134
20
121
15
121
30
126
12
105
25
121
30
121
15
134
4
134ºC
0.7 bar 5+ 10'
Tel 34 937 700 877 Fax 34 937 702 362
Pag.: 36
Without
sensor
105-130ºC
115-130ºC
115-130ºC
105-130ºC
Vacuum
Time
purge
fractions *
10
2
30
3
20
3
15
2
30
3
25
3
15
1
-
1
30
1
30
1
3' 30''