As a safety feature, this product does not operate if the Foot Control is depressed when the Chamber is mounted incorrectly. An
alarm sounds heard and the tooth icon and "POWDER" flash alternately. Mount the Chamber correctly.
CAUTION
The Perio-Mate Powder (option) scatters easily because it is made of very fine particles. Fill the Chamber slowly to prevent the
powder from scattering, and be careful not to swallow the powder or allow it to get into eyes.
FLASH pearl is almost insoluble in water. To avoid clogging of the pipes by using FLASH pearl, periodical cleaning and
maintenance of the vacuum hose, saliva ejector and separator are necessary.
8-3 Attaching/removing the Nozzle Tip
8-3-1 Attaching the Nozzle Tip
Attach the Nozzle Tip (option) securely and firmly to the Nozzle of the Powder
Handpiece (for Perio use) (Fig. 27).
Do not use a deformed or damaged Nozzle Tip. It may detach during
use.
WARNING
Be sure to attach the Nozzle Tip securely and firmly as illustrated
(Fig. 28). If the Tip is attached incorrectly, it will fall off and may be
CAUTION
accidentally ingested by the patient.
When attaching a Nozzle Tip to the Powder Handpiece, always wear
sterile surgical gloves and make sure that the Powder Handpiece is
sterilized.
8-3-2 Removing the Nozzle Tip
Use the Perio-Mate Nozzle Tip Remover (option) to remove the Nozzle Tip (Fig. 29)
and dispose of the Nozzle Tip when removed.
The Nozzle Tip can only be used once. Do not reuse or sterilize the
Nozzle Tip as it may be damaged or contaminated. If the Nozzle Tip is
WARNING
reused, it may not operate correctly due to deformation during removal,
or it may detach during use resulting in accidental ingestion by the
patient. The same applies to a Tip that has been accidentally removed.
8-4 Power Level Setting
Touch the Power Level Keys ( / ) to set the Power
Level. The Power Level Display and the Power Level
Indicator increase or decrease in accordance with
the selected Power Level (Fig. 30).
NOTICE
If you specify this setting without
attaching the Chamber, the Power Level
when using the Prophy mode is set.
8-5 Irrigation setting
1) Touch the Irrigation Volume Keys ( / ) to adjust the Irrigation Volume.
2) Depress the Foot Control to spray powder into a spittoon, etc., to confirm that powder, air and water are appropriately sprayed.
3) Adjust the Irrigation Volume while checking the irrigation from the Nozzle.
NOTICE
If the Irrigation Volume is too low with the Irrigation Volume set to 10 or the Irrigation Volume is too high with the Irrigation
Volume set to 1, adjust the basic Irrigation Level. Refer to "11-3 Basic Water Level Setting" for details.
If you perform adjustment without attaching the Chamber, the Irrigation Volume when using the Prophy mode is set.