.CALIBRATION
This scale has been calibrated at the factory and does not require re-calibration prior to use.
The calibration procedure may be performed using a known weight in either kilograms or pounds. Simply set the scale for the desired unit of
measure using the KG/LB key and start the calibration procedure. The following instructions assume the scale is set to measure lbs. The
calibration is performed using kg or lb, according to the unit of measure used upon entering into the calibration path. Note: The calibration
weight must be no less than 60 lb (28 kg) and no more than 300 lb (135 kg).
The calibration is performed using kg or lb, according to the unit of measure used upon entering into the calibration path.
OPERATOR ACTION
While display module is turned on and "START' is displayed on the LCD
(for 3 seconds); Press ZERO-KG/LB-KG/LB-ZERO
Press ENTER
Press ENTER to calibrate with 300lb
Press BMI to change the weight and then the arrows to change the
flashing digit, BMI to move to the next digit, until you reach the
required weight (no less than 60lb and no more than 300lb)
Clear the platform and press ENTER
Wait without touching or shaking the scale until the zero calibration
process is finished
Load the platform with the required weight and press ENTER
Press ENTER
Remove weight from the platform
Wait until the scale resumes normal operation
.EXPLODED VIEW OF
PROCEDURE.
"CAL" will appear on the LCD
Enter load weight "Load/300.0"
"CLEAR"
Zero Calibration
"-----------"
Weight calibration
"Put XXX.X"
Weight calibration
"Put XXX.X"
"DONE"/"CLEAR"/XXX
Rebooting, please wait
"0.0"
SCALE.
DISPLAY
6