12. Battery Replacement

This device is supplied with 2 new, long-life 1.5V, size AAA
batteries. Batteries need replacing when this icon «» BO is the
only symbol shown on the display.
Remove the battery cover BR by sliding it in the direction shown.
Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
Batteries and electronic devices must be disposed of in
accordance with the locally applicable regulations, not with
domestic waste.

13. Guarantee

This device is covered by a 5 year guarantee from the date of
purchase. The guarantee is valid only on presentation of the guar-
antee card completed by the dealer (see back) confirming date of
purchase or the receipt.
• The guarantee covers the device. Batteries and packaging are
not included.
• Opening or altering the device invalidates the guarantee.
• The guarantee does not cover damage caused by improper
handling, discharged batteries, accidents or non-compliance
with the operating instructions.
Please contact Microlife-service.

14. Technical Specifications

Non Contact Thermometer NC 150
Type:
Body mode: 34.0-42.2 °C / 93.2-108.0 °F
Measurement
Object mode: 0-100.0 °C / 32-212.0 °F
range:
0.1 °C / °F
Resolution:
Laboratory:
Measurement
±0.2 °C, 36.0 ∼ 39.0 °C / ±0.4 °F, 96.8 ∼ 102.2 °F
accuracy:
Liquid Crystal Display, 4 digits plus special icons
Display:
The unit is turned ON and ready for the measure-
Acoustic:
ment: 1 short beep.
Complete the measurement: 1 long beep (1 sec.) if
the reading is less than 37.5 °C / 99.5 °F, 10 short
«beep» sounds, if the reading is equal to or greater
than 37.5 °C / 99.5 °F.
System error or malfunction: 3 short «bi» sounds.
30 readings recall in the memory mode with a
Memory:
record of both time and date.
NC 150
Backlight:
The display light will be GREEN for 4 seconds,
when the unit is turned ON.
The display light will be GREEN for 5 seconds,
when a measurement is completed with a reading
less than 37.5 °C / 99.5 °F.
The display light will be RED for 5 seconds, when
a measurement is completed with a reading equal
to or higher than 37.5 °C / 99.5 °F.
Body mode: 16-40.0 °C / 60.8-104.0 °F
Operating
Object mode: 5-40.0 °C / 41-104.0 °F
conditions:
-20 - +50 °C / -4 - +122 °F
Storage
15-95 % relative maximum humidity
conditions:
Approx. 1 minute after last measurement has been
Automatic
taken.
Switch-off:
2 x 1.5 V alkaline batteries; size AAA
Battery:
approx. 2000 measurements
Battery
(using new batteries)
lifetime:
Dimensions: 141.1 x 43.3 x 36.9 mm
90 g (with batteries), 67 g (w/o batteries)
Weight:
IP21
IP Class:
ASTM E1965; IEC 60601-1; IEC 60601-1-2 (EMC);
Reference to
IEC 60601-1-11
standards:
Expected
service life: 5 years or 12000 measurements
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
According to the Medical Product User Act a biennial technical
inspection is recommended for professional users. Please
observe the applicable disposal regulations.

15. www.microlife.com

Detailed user information about our thermometers and blood pressure
monitors as well as services can be found at www.microlife.com.