Código de error
Error Code
EE-Error
FR-Error
LV-Error
80
Length
7 characters ASCII
6 characters ASCII
Length
4 characters ASCII
5 characters ASCII
Length
4 characters ASCII
7 characters ASCII
Name
Error While Storing Data in EEPROM
Erorr in Frame (too short, too long, wrong data)
Value out of limits
VEGABAR 82 • Protocolo Modbus y Levelmaster
Code/Data
UuuRmmm
mmm = milliseconds (50 up to 250),
default = 127 ms
UuuROK
Code/Data
UuuF
UuuFn
n = number of measurement values
(0, 1 or 2)
Code/Data
UuuR
UuuRmmm
mmm = milliseconds (50 up to
250), default = 127 ms