Installation; Setting Up Light Source; Connecting Light Cable - Cuda surgical ELITE 300 Manual De Uso

Ocultar thumbs Ver también para ELITE 300:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

5. INSTALLATION

5.1 SETTING UP THE LIGHT SOURCE
Place the light source on a stable surface (cart, counter, stand, etc.).
NOTE
Avoid places where the light source may be splashed with
liquid.
Absolutely DO NOT use in any environment with explosive or
flammable gases.
DO NOT block the venting grids of the light source.
Make sure the power switch is in OFF position.
Connect AC power cord to the power inlet located on the rear panel of the light source.
Make sure power cord is completely seated into the power inlet.
CAUTION Use only cords provided with the light source.
CAUTION To prevent electric shock, connect power cords of peripheral
equipment through medical isolation transformers.
NOTE
When using medical isolation transformer, be sure to check the
transformer power ratings.
Make sure that the power cord is connected to the main power
with three prong plug (USA use UL2601-1 rated isolation
transformers and/or power strips only).
CAUTION When used in clinical or residential areas near radio or TV units,
this equipment may be subjected to radio interference. To avoid
adverse electromagnetic effects, DO NOT operate this equipment
near RF energy equipment.
5.2 CONNECTING THE LIGHT CABLE
Connect the light cable to endoscope and then plug the light cable endfitting into the
appropriate port of the turret on the front panel.
LIT037 CUDA
®
SURGICAL ELITE 300
(English)
1 2
3 4
Rev. L
Page 8 of 60

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Lit037

Tabla de contenido