Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The SwiSto3 is intended for use in the electromagnetic environment specified below.
The customer or user is to assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations on
power-supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
Note: UT is the ac mains voltage prior to application of the test level.
IEC 60601
test level
± 6 kV contact discharge
± 8 kV air discharge
±2 kV for power supply
lines
±1 kV for input/output
lines
± 1kV voltage
line to line
± 2kV voltage line
to ground
<5 % UT
for 1/2 cycle
(>95 % dip)
40 % UT
for 5 cycles
(60 % dip)
70 % UT
for 25 cycles
(30 % dip)
<5 % UT for
5 seconds
(>95 % dip)
3 A/m
Compliance level
± 6 kV contact discharge
± 8 kV air discharge
±2 kV for power supply
lines
±1 kV for input/output
lines
± 1kV voltage
line to line
± 2kV voltage line
to ground
<5 % UT
for 1/2 cycle
(>95 % dip)
40 % UT
for 5 cycles
(60 % dip)
70 % UT
for 25 cycles
(30 % dip)
<5 % UT for
5 seconds
(>95 % dip)
3 A/m
27
Electromagnetic environment –
guidance
Floors should be wood, concrete,
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the device
requires continued OPERATION during
a power mains interruption, it is
recommended that the device be
powered by an uninterruptible power
supply or battery.
If interference occurs, it may be
necessary to place the SwiSto3
further away from the sources of
power frequency magnetic fields or to
install a magnetic shield: The power
frequency magnetic field should be
measured at the designated site to
ensure that it is not large enough to
cause interference.
ENGLISH
www.kawemed.com