Maintenance

Maintenance
The unit may only be serviced and repaired by service
personnel employed or authorized by Dräger Medical.
The unit must be inspected and serviced regularly at yearly
intervals or after 5000 hours of operation (and a record kept).
Correct functioning must be verified by service experts every
6 months.
A service manual is available to each authorized service expert
for service, maintenance and repair purposes.
The unit may only be opened by authorized service
experts in order to avoid damage due to incorrect
maintenance.
Always unplug the mains cable before opening the unit in
order to avoid the risk of electric shock.
Only use original Dräger Medical parts when servicing the
unit, otherwise faults or defects may arise.

Maintenance intervals

Required action
Inspection and maintenance
Correct functioning
Dust filter
Internal intake filter
Internal bacterial filter
Compressor
All inhalation tubes and filters

Disposal of the unit

At the end of its service life:
Permox SilentCare must be disposed of as required in
consultation with the relevant waste management companies.
The applicable regulations must be observed.
This "Instructions for Use" manual is published for public use and is only meant for your information.
We do not guarantee the accuracy of these instructions with respect to your specific device type or status.
Only the "Instructions for Use" manual delivered together with a device is officially published for the use of that specific device.
24
Intervals
By service experts once per year or after 5000 hours of operation.
Must be checked by service experts every six months.
Must be replaced every month or when clogged.
Old filters can be disposed of as household refuse.
Must be replaced by service experts every six months.
Must be replaced by service experts once per year or after 5000 hours of
operation.
Must be serviced by service experts every 3 years or after 12000 hours of
operation.
Must be replaced for each new user or as directed by the doctor.