Fault – Cause – Remedy
If a fault arises, follow the recommended step-by-step procedure.
Fault
Unit does not run,
continuous acoustic alarm
Unit runs,
red alarm lamp flashes,
intermittent acoustic alarm
Unit does not run,
red lamp lights up constantly,
continuous acoustic alarm
Unit runs,
no acoustic alarm,
no alarm lamp lights up,
no oxygen supply
This "Instructions for Use" manual is published for public use and is only meant for your information.
We do not guarantee the accuracy of these instructions with respect to your specific device type or status.
Only the "Instructions for Use" manual delivered together with a device is officially published for the use of that specific device.
Cause
Unit has been switched on but the mains
cable is not plugged in.
Fuse has blown.
Oxygen concentration is too low because
a hose has been kinked or is not
connected correctly.
Oxygen concentration is too low because
a hose is fouled.
Too little air is taken in because the dust
filter is fouled.
Oxygen flow has been changed several
times over the whole flow range within a
very short space of time.
Device error
No oxygen supply because a hose has
not been connected.
The unit has overheated because the
dust filter is fouled.
The unit may have overheated or an
internal fault may have occurred because
the oxygen flow has repeatedly been
changed over the entire range within a
very short space of time.
Device error
Oxygen is not supplied because a hose
has not been connected.
Fault – Cause – Remedy
Remedy
1. Check that the mains cable
has been connected correctly.
2. Call a service expert.
1. Ensure that all hoses have been
connected correctly and are not
kinked.
2. Ensure that all hoses
are clean.
3. Clean the dust filter, page 23.
4. Switch off the unit, leave it to cool for
approx. 30 minutes and switch on
again. Do not change the set flow.
5. Call a service expert. The unit can still
be used without reservation, but the
oxygen concentration may be too low.
1. Ensure that all hoses have been
connected correctly.
2. Clean the dust filter, page 23.
3. Switch off unit, leave it to cool for
approx. 30 minutes and switch on
again. Do not change the set flow.
4. Call a service expert.
1. Ensure that all hoses and the
humidifier unit have been connected
correctly.
25