Clean The Instrument; Clean Spills; Errors-Red Light; Warnings-Yellow Light - Hach CL17sc Manual Del Usuario

Ocultar thumbs Ver también para CL17sc:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 112
7. Remove the pump clamp. Refer to
8. Remove the tubing harness and stir bar from the analyzer. Keep the tubing harness for shipment
with the analyzer. Refer to
9. Install the pump clamp without the tubing harness. Put tape on the tubing clamp to hold the tubing
clamp tightly.
10. Remove the water from the cell with a disposable dropper or a no-lint cloth.
11. Remove the analyzer from the wall.
12. Put the analyzer back in the original packaging.

7.7 Clean the instrument

Never use cleaning agents such as turpentine, acetone or similar products to clean the instrument
including the display and accessories.
Clean the exterior of the instrument with a moist cloth and a mild soap solution.

7.8 Clean spills

Chemical exposure hazard. Dispose of chemicals and wastes in accordance with local,
regional and national regulations.
1. Obey all facility safety protocols for spill control.
2. Discard the waste according to applicable regulations.
Section 8 Troubleshooting
8.1 Errors—Red light
To show the errors, push menu and select DIAGNOSTICS > [select analyzer] > ERROR LIST. A list
of possible errors is shown in
The analyzer status light changes to red when an error occurs. When an error occurs, measurements
stop, the measurement screen flashes and all outputs are held as specified in the controller menu.
Error
DETECTOR ERROR A detector does not operate correctly.
DIRTY CELL
CODE ERROR
LED ERROR
PUMP ERROR
SAMPLE LEAK
8.2 Warnings—Yellow light
To show the warnings, push menu and select DIAGNOSTICS > [select analyzer] > WARNING LIST.
A list of possible warnings is shown in
Figure 6
Step 3: Install the stir bar and tubing harness
N O T I C E
C A U T I O N
Table
7.
Table 7 Error messages
Description
The cell is stained or dirty.
A firmware error occurred.
The light in the cell does not operate
correctly.
The pump does not operate correctly.
There is a sample leak in the analyzer.
Table
8.
on page 24.
Contact technical support.
Clean the cell. Refer to
the cell
Contact technical support.
Contact technical support.
Contact technical support.
Contact technical support.
on page 13.
Solution
Clean
on page 20.
English 25

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido