7) Remove the probe from the control unit.
Never hold the probe cord when removing the probe from the
control unit. Be sure to hold the connector part.
CAUTION
The root canal length cannot be correctly measured in some cases.
ALWAYS confirm in combination with X-ray photography.
Example:
1. The apical foramen is large.
2. Closed root canal.
3. Bleeding from the root canal orifice.
4. Tooth crown is broken.
5. Tooth root is broken.
6. The root canal is filled with gutta-percha.
7. Metal prosthesis on the tooth crown is in contact with gum.
8. Immature tooth or broken root tooth.
9. The root canal is extremely dry.
7

Operation check using the Tester

If the numerical value is not displayed correctly or the value flashes, follow the
steps below to test and verify operation.
1) Press the Power Key to turn ON the
power. (The alarm sounds and the
LCD panel lights up.)
2) Pinch one terminal on the Tester with
the file clip and contact the lip hook to
the other terminal. (Fig.7)
3) Check that the number display on the
LCD panel is within the range of "0.4
and 0.6". If not correctly displayed,
follow the step 4.
4) Remove the file clip and lip hook from the probe then insert the probe sockets
directly into the Tester (Fig.2). Check that the number display on the LCD
panel is within the range of "0.4 and 0.6". If not correctly displayed, follow
the step 5.
Fig. 7