• MR suitability of the seca b11 and b12
The seca b11 and b12 pressure gages were tested for suitability for use in a
magnetic resonance (MR) environment in combination with the cuffs and their
Velcro closures. Both the requirements on functionality and on electromagnetic
attractive force were met with a range using 3 Tesla. This is indicated by the
marking "MR Conditional" on the scale.

4 . HYGIENE TREATMENT

4 .1 Cleaning and caring for the pressure gage
NOTICE! Damage to device
Unsuitable cleaning agents and disinfectants may damage the sensitive
surfaces of the device.
• Do not use aggressive or abrasive cleaning agents.
• Do not use white spirit or petroleum spirit.
• Only use disinfectants suitable for sensitive surfaces. Suitable disinfectants
are available from specialist dealers.
• For cleaning, we recommend a damp cotton or microfiber cloth wetted
with disinfectant if appropriate.
• Prevent liquids from penetrating the housing.
4.2 Cleaning and caring for the cuffs
Care information for seca blood pressure cuffs with Velcro closure and D-ring:
NOTICE! Damage to device
Never autoclave the cuffs.
Washing
Once the bladder has been removed, at up to 30 °C + detergent for delicates.
Fold covers with Velcro closures before washing so that the two halves of the
Velcro tape are in contact.
Cleaning by
We recommend the following cleaning products and disinfectants:
spraying or
wiping with
disinfectant
Following cleaning/disinfecting, wipe over with clean water to remove all residues
and then leave to air-dry. Please use only soft cotton or microfiber cloths.
• mild soap
• Bacillol AF
• Meliseptol
• Oxivir Plus spray
• Terralin Protect
• 70%-strength isopropyl alcohol
• Fugaten
spray (Lysoform spray)
®
English
• 9