Sample preparation

22
25 mg/l, as otherwise the electrode could be damaged irreversibly. If
you expect the concentration to be higher, dilute the sample according-
ly.
In the case of an erroneous calibration immediately after commissioning, pol-
ish the membrane surface (see M
Malfunctions can possibly occur due to the use of a magnetic stirrer. If the val-
ues are fluctuating heavily, we recommend to switch off the magnetic stirrer
during measurement.
Sample preparation
Add 1 % MZ/NH3/CN solution. This sample conditioning solution creates
optimum conditions for measuring. It provides a constant ionic strength and
similar diffusion potentials at the reference electrode in standard solution and
test sample.
If you would like to have more detailed information concerning sample prep-
aration and measuring procedures, WTW provides a large number of appli-
cation reports for various applications.

Response times

The response time depends on the concentration range. It is
 several seconds at high concentrations,
 several minutes near the detection limit.
The measured value is stable if the value does not change by more than
0.1 mV within 30 seconds.

Interferences

 Complexes with Ni
 Strong reducing agents
 Interfering ions: 10 % error with the following concentration ratio
(concentration ratio = interfering ion / measured ion):
-
-
Cl
I
6
1x10
0.1
AINTENANCE
2+
2+
2+
, Cd
, Cu
-
2-
Br
S
3
5x10
must not be present
CN 500 / CN 800
).
ba15356e03
03/2015