Sample Weight And Volume (Sample Quantity) - Xylem SI Analytics TitroLine 5000 Manual De Instrucciones

Ocultar thumbs Ver también para SI Analytics TitroLine 5000:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

118
Fig. 71

4.6.3.2 Sample weight and volume (sample quantity)

The Sample Quantity (W) item (Fig. 72) is used to select whether one is wishing to use a sample weight or a
sample volume for titration or solution preparation (Fig. 73).
Fig. 72
Fig. 73
You have the following options:
«Manual sample weight»: The sample weight is enquired by a prompt at the start of the method and
manually input.
«Automatic sample weight»: The sample weight is automatically transferred by a connected balance.
«Fixed sample weight»: A fixed sample weight is input in g. This weight will then automatically be used
for each start of the method.
«Manual sample volume»: The sample volume in ml is prompted at the start of the method and
manually input.
«Fixed sample volume»: A fixed sample volume is input in ml. This volume will then automatically be
used for each test of the method.

Publicidad

Tabla de contenido
loading

Tabla de contenido