TROUBLESHOOTING

During operation, any rattling or ticking sounds may indicate a loose screw on the platform tray, tray attachment or accessory. All accessories should be sufficiently tightened in
place before starting the unit.
Error
Cause of Error
E01
RTD open or temperature over 100°C (212°F)
E02
RTD shorted or temperature below 0°C (32°F)
ceased bearing
drive belt broken
E03
mechanical obstruction
loose foot (suction cup)*
maximum load exceeded
E04
loose foot (suction cup)*
9
How to Fix
This error should NOT be addressed by the end user. Switch the unit off and contact your Ohaus
representative for repairs.
This error should NOT be addressed by the end user. Switch the unit off and contact your Ohaus
representative for repairs.
Press the standby button to clear this error and remove the mechanical obstruction. If the E03 error
persists the reason may be a ceased bearing or broken drive belt and should NOT be addressed by
the end user. Switch the unit off and contact your Ohaus representative for repairs.
* In the event a foot (suction cup) has come loose from the bench top, the unit will register an errant
E04 or E03 error message due to the instability of the unit. Press the standby button to clear this error.
Firmly press down on the four (4) corners of the unit, creating a strong suction to the work surface (DO
NOT place on bench mat). Press the standby button to resume operation.
Press the standby button to clear this error. Be sure the load is within the maximum load capacity before
restarting the unit. If the E04 error persists, switch the unit off and contact your Ohaus representative
for repairs.
* In the event a foot (suction cup) has come loose from the bench top, the unit will register an errant
E04 or E03 error message due to the instability of the unit. Press the standby button to clear this error.
Firmly press down on the four (4) corners of the unit, creating a strong suction to the work surface (DO
NOT place on bench mat). Press the standby button to resume operation.