Troubleshooting - Philips Respironics InnoSpire Mini Guia De Inicio Rapido

Compresor
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Troubleshooting

Problems
Solutions
Compressor overheats. The
The thermal protector has been activated due to:
compressor has an internal
a. The device operating outside the conditions outlined in this
thermal protector to prevent
manual.
the device from overheating.
b. The device has been operating near heat sources or in hot
If the internal temperature
environments.
reaches the preset level, the
Allow the compressor to cool down for a minimum of
unit will automatically shut
40 minutes. Wait until the red light turns off before turning
down, and the ON/OFF
the compressor on again.
button light will turn red.
The compressor does not
Check all power sources to the compressor. Push the ON/OFF
operate.
button again.
Medication does not nebulize
Check for proper connection of the tubing adapter.
properly.
Make sure the nebulizer is not empty.
Clean the nebulizer as described in the manual.
Replace the nebulizer.
Check the intake filter and replace it if it is dirty.
The device is noisy.
Check that the filter cover is correctly positioned.
The compressor does not
Make sure battery contacts cover is removed
operate when connected
Check the assembly of the battery pack to compressor
to battery.
Compressor is disabled when battery is charging. Remove
charger from battery to operate the compressor.
Ensure battery pack has adequate charge. Recharge battery.
Short circuit protection may have been activated. This fault
condition can be re-set by charging the battery pack.
Should the device still not operate properly after checking the unit as indicated above, contact
Philips Respironics Customer Service at 1-800-345-6443 or 1-724-387-4000.
12
Specifications
Power supply
See adapter rating label
Weight
0.93 lbs.
Size
4.5" x 3.5" x 2.1" (114.3 mm x 88.9 mm x 53.3 mm)
Noise level
50
+
/
_
Average flow rate
3.4 LPM @12.7 psi
Maximum pressure
30 psi
Maximum flow rate
~ 8 LPM
Mode of operation
Intermittent use 30 min ON / 30 min OFF
Class II device (double safety insulation).
Type BF device (device with specific protection against electrical hazards).
Certification
Electric safety standards EN 60601-1
Electromagnetic compatibility according to EN 60601-1-2
Storage conditions
Temperature
MIN -4° F MAX 140° F (MIN -20° C MAX +60° C)
Humidity
MIN 10% RH –MAX 95% RH
Atmospheric pressure
Performance may vary depending on altitude above sea level,
barometric pressure and temperature
Operating conditions
Temperature
MIN 41° F MAX 104° F (MIN +5° C MAX +40° C)
Humidity
MIN 10% RH –MAX 95% RH
Atmospheric pressure
Performance may vary depending on altitude above sea level,
barometric pressure and temperature
Expected service life
Compressor 3 years from date of purchase
Battery 6 months from date of purchase
Replacement parts and accessories
Filter
1109488
AC power/battery charger
1109441
12V DC (car) adapter
1109455
Rechargeable lithium ion battery
1110726
ENGLISH