Figure 6 Multiple bottle installation
1 Retainer for 24 1-L poly bottles
2 Retainer for 24 350-mL glass
bottles
3 Bottle one indicator
Plumb the sampler
Install the intake tube in the middle of the sample stream (not near the surface or bottom) to make
sure that a representative sample is collected.
1. For a sampler with the standard liquid detector, connect the tubing to the sampler as shown in
Figure
7.
Note: When Teflon-lined tubing is used, use the tubing connection kit for Teflon-lined PE tubing.
2. For a sampler with the optional non-contacting liquid detector, connect the tubing to the sampler
as shown in
Figure
Note: When Teflon-lined tubing is used, use the tubing connection kit for Teflon-lined PE tubing.
3. Install the intake tube and strainer in the main stream of the sample source where the water is
turbulent and well-mixed. Refer to
• Make the intake tube as short as possible. Refer to
intake tubing length.
4 Bottle tray for 8 to 24 bottles
5 Slot for bottle tray
6 Floor of refrigerated sampler
8.
Figure
9.
Specifications
7 Front of sampler
8 Removable tray (AWRS only)
9 Insert (refrigerated sampler only)
on page 3 for the minimum