Specifications
Specifications are subject to change without notice.
Refrigerated Sampler and All Weather Refrigerated Sampler (AWRS)
Dimensions (W x D x
1
H)
Weight
Power requirements
(includes compressor)
Overload protection
Compressor
Operating temperature
Storage temperature
Relative humidity
Installation category,
pollution degree
Protection class
Temperature control
1
Refer to
2
Radio frequency interference in the 30 to 50 MHz range can cause a maximum temperature
change of 1.3 °C (34.3 °F). Adjust the set point temperature to 2 to 10 °C (35.6 to 50 °F) to
correct this interference.
Refrigerated Sampler
61 x 61 x 112 cm (24 x 24 x 44 in.)
63.3 kg (140 lb)
115 VAC, 60 Hz, 3.3 A (18 locked
rotor amps)
230 VAC, 50 Hz, 1.7 A (9 locked rotor
amps)
Thermal overload relay, opens at
100 to 110 °C (212 to 230 °F)
1
/
HP
6
0 to 50 °C (32 to 122 °F)
-30 to 60 °C (-22 to 140 °F)
0 to 95%
II, 2
I
4 (±0.8) °C (39 (±1.5) °F) in ambient
temperatures at a maximum of 50 °C
(120 °F)
on page 6 for the sampler dimensions.
Startup
on page 18
Maintenance
on page 19
Troubleshooting
on page 25
AWRS
76 x 81 x 130 cm (30 x 32 x 51 in.)
86 kg (190 lb)
115 VAC, 60 Hz, 4.2 A or 6.4 A with controller
compartment heater
230 VAC, 50 Hz, 2.7 A or 4.1 A with controller
compartment heater
115 VAC: 7.5 A circuit breaker
230 VAC: 5.0 A circuit breaker
1
/
HP
5
115 VAC: 115 °C (239 °F) thermal overload
protector, 7.1 locked rotor amps
230 VAC: 120 °C (248 °F) thermal overload
protector, 7.6 A peak start current
0 to 50 °C (32 to 122 °F); with AC battery
backup: 0 to 40 ºC (32 to 104 ºF); with
controller compartment heater: -40 to 50 °C
(-40 to 122 °F); with controller compartment
heater and AC battery backup: -15 to 40 ºC
(5 to 104 ºF)
-30 to 60 °C (-22 to 140 °F)
0 to 95%
II, 2
I
4 (±0.8) °C (39 (±1.5) °F)
2
English 3